DDT-320-000 Rev G Belmont Medical Technologies
Page
79
of
114
6. For residue removal, use a new lint free cloth moistened with sterile
water. Use the cloth on the exterior of the system, the screen, and the
hoses.
Before Storage
1. Use PPE as recommended by the disinfectant’s manufacture.
2. With the device in Standby Mode, close the clamps on the wrap.
3. Remove the wrap from the patient; disconnect it from the hoses
and dispose of it.
4. Disconnect the hoses from the machine.
5. Dispose of disposable temperature probes and the disposable
adapter in accordance with hospital procedures for medical waste.
Disinfect the reusable temperature probes as required by the
manufacturer’s directions.
6. Insert the volume of Sodium Dichloroisocyanurate (NaDCC) powder
or tablets as recommended by NaDCC manufacturer into 1.2 liter
water tank.
7. Run the device in Standby Mode for 30 minutes.
8. Empty the device (see
Figure 27: Select Service
).
9. Turn off the machine. Wait a few seconds. Unplug the power cord.
Thermal Disinfection Process (Self-Cleaning)
This feature performs a thermal disinfection of the water tank and internal
tubing.
The thermal disinfection of CritiCool® MINI is an integrated feature, which
heats the circulating water of the system, thus allowing the heat to
disinfect the internal water pathways of the system, including the water
tank.
Thermal disinfection is performed at every Periodic Maintenance and can
only be performed by a Belmont certified technician.
Required Equipment
•
Bypass tube PN #200-00181 or PN #200-00096
•
Up to 1.2 liters of sterile or 0.22
μ
m filtered water
To perform Thermal Disinfection:
Make sure that water tank is full, and the bypass tube is connected.