DDT-320-000 Rev G Belmont Medical Technologies
Page
86
of
114
CAUTION!
The repair and servicing of the CritiCool
®
MINI system
should be performed only by Belmont Medical
Technologies or authorized agents of Belmont Medical
Technologies.
Figure 42: CritiCool MINI System Malfunction (No Message) Troubleshooting Guide
Observation
Possible Problem
Action to be Taken
The power switch of
CritiCool
®
MINI is set to "ON"
but it is not activated, and the
control panel is blank.
CritiCool
®
MINI is
unplugged.
Check the power cable
connections.
No line voltage
Call Biomedical Department.
Wrap begins to leak.
The wrap was
accidentally punctured
during the course of the
operation.
Turn off the CritiCool
®
MINI
and allow the water to return
to the reservoir.
Replace the Wrap if possible.
Water leaks from the
connector between Wrap and
the connecting tube.
Connecting tubes are
not sealed properly.
Close clamps on Wrap.
Disconnect connecting tubes
and re-connect connecting
tubes until the click sound is
heard.
Damage to connecting
tubes.
Replace connecting tubes.
Damage to quick coupling
connector.
Call Biomedical Department.
Water leaks between
connecting tubes and CritiCool
®
MINI.
Connecting tubes are
not connected
properly.
Disconnect connecting tubes
from the machine and re-
connect again until the click
sound is heard.
Damage to connecting
tubes.
Replace connecting tubes.
Damage to quick
coupling connector.
Call Biomedical Department.