An audible alarm also sounds.
Touching the hard key next to the Alarm icon
will mute the alarm
for
five
minutes but leave the message on the screen.
NOTE:
If the temperature is below 30.5°C, the alarm cannot be silenced.
NOTE:
If the user disregards the message and does not touch OK for over
30 minutes, the alarm cannot be silenced
.
While the message appears, thermoregulation is paused and the machine
switches into standby mode (water stops flowing to the wrap).
Check that the core temperature probe is in place and the low
temperature represents the true patient status and then touch
OK
to re-
activate temperature control.
When
OK
is touched, the screen returns to the main screen and the
following message appears for 5 seconds.
Figure 32: Temperature Regulation Paused –
Core Readout Too Low Message