DDT-320-000 Rev G Belmont Medical Technologies
Page
62
of
114
Message
Screen Message
Description
Patient temperature is
above 38.5°C
An alarm sounds but
thermoregulation continues.
The alarm can be muted for
30 minutes.
Low Core Temperature
This message appears
when core temperature is
>0.8°C less than the set
point or according to
alarm settings.
Patient's Temperature
is below XX.X°C
An alarm sounds but
thermoregulation continues.
NOTE:
It is possible to change the range of these alarms in the
Settings screen. The user can choose at which temperatures
the "High Patient Temp" and "Low Patient Temp" alarms will
be activated.