Mode Select
The MODE SELECT panel enables choosing the mode of operation:
•
TTM (Targeted Temperature Management)
Use this mode for Targeted Temperature Management. This mode
is useful for any procedure where thermoregulation is required to
bring the patient's temperature to a stable set point temperature.
•
Controlled Rewarming
This mode provides gradual controlled rewarming. Each step of the
procedure increases the set point temperature by a fixed, small step
in temperature for a predefined period. The step is always related to
the core temperature reached at the end of the previous stage. From
the settings screen, you can choose the rewarming rate.
•
•
•
•
Normothermia
This mode is for a fast warming in cases in which a patient
needs to be rewarmed quickly. This mode should not be used
for patients undertaking cooling therapy.
NOTE:
When switching to Normothermia, the system keeps the last set
point of the preceding mode.
To select a mode:
1. Touch the MENU icon
.
2. Touch
Mode Select
to display the select mode panel.
3. Touch the required mode icon. The selected mode is now highlighted.
4. Touch
OK t
o activate the mode.
Figure 4-7: Select Mode panel