Overview and Installation
Introduction
The purpose of the CliniLogger™ device is to save the Allon® /
CritiCool® systems' vital data for further reference. By means of the
CliniLogger™ Viewer software, the user can use an external PC to
review this saved data.
Using the CliniLogger™ Application
The CliniLogger™ device connects to the RS-232 (serial) connector in
the rear of the Allon
®
for data transfer. While the device is connected,
data is saved at one-minute intervals.
Connect the CliniLogger™ device to the Allon before the start of the
medical procedure.
Belmont Medical Technologies recommends recording Allon® device
data for one patient at a time. At the end of the procedure, disconnect the
CliniLogger™ device from the thermoregulation machine and connect to
a PC. Download the data from the device and then reconnect the
CliniLogger™ to the Thermoregulation machine so it is ready for the next
procedure.
The CliniLogger™ Software
The CliniLogger™ device is supplied with a CliniLogger™ Viewer
software CD to be installed on a PC for downloading and viewing the
saved data from the Allon.
Installing the Software
To install the CliniLogger™ software:
1. On your PC, double-click on
My Computer
and open the CD drive.
2. Double-click the
Installer
folder.
3. Double-click the
Volume
folder.
4. Double-click
setup
; the CliniLogger™ install window appears.