WARNING!!!
The repair and servicing of the Allon
®
system
should be performed only by Belmont Medical
Technologies or authorized agents of Belmont
Medical Technologies.
Table 10: Allon
®
System (No Message) Troubleshooting Guide
Observation
Possible Problem
Action to be Taken
The power switch of the Allon
®
device is set to "ON" but it is not
activated and the control panel is
blank.
Allon®
device is unplugged.
Check the 115/230 VAC power cable
connections.
No line voltage
Call the biomedical technician.
ThermoWrap
®
begins to leak.
The ThermoWrap
®
was
accidentally punctured during
the course of the operation.
Turn off the Allon
®
device and allow
the water to return to the reservoir.
Replace the ThermoWrap
®
if possible
Water leaks from the connector
between ThermoWrap
®
and the
connecting tube.
Connecting tubes are not
connected properly.
Close clamps on ThermoWrap
®
.
Disconnect connecting tubes and re-
connect connecting tubes until the
click sound is heard.
Damage to connecting tubes.
Replace connecting tubes.
Damage to Quick Coupling
Connector.
Call the biomedical technician.
Water leaks between connecting
tubes and the Allon
®
device.
Connecting tubes are not
connected properly.
Disconnect connecting tubes from
the machine and reconnect again.
Damage to connecting tubes.
Replace connecting tubes.
Damage to Quick coupling
connector.
Call the biomedical technician.