Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies
Page
38
of
106
Figure 13: Control Panel
Getting Started
Preparing the System for Operation
To prepare the system for operation:
1.
Place the unit in the desired position according to Space and
Environmental Requirements in Chapter 3.
CAUTION!
Do not place the Allon
®
device under the operating
table or the patient’s bed.
2.
Press the brake pedals and lock the wheels to secure the Allon® device.
3.
Remove the water tank feeder cover and pour in sterile water or
0.22µ filtered water until the maximum allowable level is reached.
CAUTION!
Do not use deionized water or water created through
reverse osmosis because it may promote corrosion of
the metal components of the system.
NOTE:
Use only sterile or 0.22 µ filtered water.
4.
Observe the water-level indicator to prevent overfilling the water
tank. Close the water tank cover.