![Belmont 200-00263 Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/belmont/200-00263/200-00263_user-manual_2730283059.webp)
Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies
Page
59
of
106
Self Cleaning
This is a feature for Belmont Medical Technologies certified technicians only. It is
password-protected.
This feature performs a thermal disinfection of the water tank and internal tubing.
The thermal disinfection of Allon
®
is an integrated feature, which heats the
circulating water of the system, thus allowing the heat to disinfect the internal
water pathways of the system, including the water tank.
Thermal disinfection is performed at every Periodic Maintenance.
CAUTION!
–
Use only sterile or 0.22 µ filtered water.
–
Do NOT use bleach or any other cleaning and disinfectant agent
for the internal circulation except sodium dichloroisocyanurate
(NaDCC). These agents may harm the system and result in
damage to the system.
–
Always drain the water after the thermal disinfection process.
NOTE:
Self Cleaning is password protected and should only be used by
authorized Belmont Medical Technologies personnel.
Thermal Disinfection Process
Required Equipment
–
Bypass tube PN #200-00181 or PN #200-00096
–
Up to 8 liters of .22 µm filtered or sterile water
To perform Thermal Disinfection:
NOTE:
Make sure that water tank is full, and that the bypass tube is
connected.
1. In the main menu, select
Services
.
2. Touch
Self Cleaning
and then
OK
.
3. The process is password protected. Insert password.