Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies Page
84
of
106
Clinical Messages and Alarms
Clinical messages call for the attention of the operator (doctor or nurse)
to the condition of the patient or call for user confirmation of the setting
by touching the OK key.
Clinical messages include the following:
Table 14: Clinical Messages and Alarms
Message
Message in window
Description
Low Core temperature
The operator needs to
confirm the location of the
core temperature probe and
touch OK to continue.
The alarm can be muted for
10 minutes.
The alarm and message are
activated according to the
selected alarm limit.
Patient’s temperature is above
Thermoregulation continues.
XX.XoC
The alarm can be muted for
10 minutes.
The alarm and message are
activated according to the
selected alarm limit.
Patient’s Temperature is below
Thermoregulation continues.
XX.XoC
The alarm can be muted for 10
minutes.
NOTE:
It is possible to change the range of these alarms in the Settings screen.
The user can choose at which temp the "High Patient Temp" and "Low Patient Temp"
alarms will be activated.