Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies
Page
61
of
106
device. Reconnect the connecting tubes to the Allon
®
device and the ThermoWrap
®
.
Reopen the clamps.
9. Turn on the Allon
®
device to resume treatment.
10. At the end of treatment, repeat steps 1–4.
11. Remove the ThermoWrap® and the temperature probes from the patient.
12. Dispose of the ThermoWrap® in accordance with hospital guidelines governing non-
toxic plastic waste.
13. Disinfect the surface of the connecting tubes and the exterior of the Allon® device (see
instructions in Chapter 6).
14. Dispose of disposable temperature probes in accordance with hospital procedures for
medical waste. Disinfect the reusable temperature probes and/or adaptor cables as
required by hospital/clinic protocol. Dispose of any damaged probes as indicated
above.
15. After each use or between cases, drop sodium dichloroisocyanurate (NaDCC) tablets
or powder into the 6.0 liter water tank and run the device for 30 minutes in Standby
Mode.
16. Store the Allon® device and its accessories in a safe place.