DDT-063-000 Rev C Belmont Medical Technologies Page
77
of
106
Table 12: Allon
®
System Messages Troubleshooting Guide
Message
Cause of Problem
Action to be
taken
Comments
Add Water
Water level is too low
Refill water tank to
maximum
The alarm can be muted for
10 minutes.
Connect Water Tubes
Connecting tubes are
not connected.
Connect
connecting tubes.
Check for snarls,
folds, or objects
that obstruct the
water flow in the
wrap.
Check clamps.
The alarm can be muted for
10 minutes.
Connect Core Sensor
No core temperature
probe is inserted in its
socket - at power up
Connect core
temperature probe
After power up, this alarm is
muted automatically for 10
minutes.
No core temperature
probe is inserted in its
socket - after power up
If the Core temperature
probe is not connected
during an operation, the
alarm can be muted for 10
minutes.