NOTE:
When there is a difference between the set point
temperature and the core temperature, a further increase in
the set point temperature will not affect the water
temperature in the ThermoWrap®. For example, if the core
temperature is 36°C (96.8°F) and the set-point temperature
is 37°C (98.6°F), raising the setpoint of the Allon® system
further will not affect the water temperature. The Allon®
device automatically operates at the optimal level to obtain
the desired set-point temperature.
NOTE:
The default setting is intended to maintain normothermia.
However, the system provides the physician with the option
of selecting a body temperature in the range of 30°C to
40°C (86°F- 104°F).
NOTE:
After the normothermia mode is selected, it takes up to 4
minutes for the system to reach equilibrium and start
adjusting the patient’s temperature according to the
programmed rewarming step. This is due to variables in
the environment: clinical, medicinal, and patient.
If the desired set point temperature is set to be out of normothermia range
(36°C- 38°C / 96.8°F-100.4°F), the message “
Out of Normothermia Range
”
appears. Touch
OK
to confirm the selected temperature.
Figure 21: “Out of Normothermia” Message