![Belmont 200-00263 Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/belmont/200-00263/200-00263_user-manual_2730283039.webp)
Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies
Page
39
of
106
NOTE:
In case of overfilling, see Table 11.
5.
Connect the Allon® device to the power source.
6.
Power on the Allon® device, which will initiate the Self-Test.
(See
Turning on the System
To turn on the system
:
1. Turn the main switch, located in the rear of the unit, upwards to the
ON position. Once power is supplied to the Allon
®
device, the unit
performs a self-test.
The self-test is performed in order to ensure the proper functioning of the
Allon
®
device. A self-test is conducted every time the system is restarted.
In the event of a power failure of less than 10 minutes, the self-test will not
be performed and the Allon
®
device will remain in operation mode.
During the self-test, the “Performing Self Test” message appears until the
system is ready.
Figure 14: Initial Self-Test Screen
Successful completion of the self-test indicates that the Allon
®
device is ready
to operate.
CAUTION!
Whenever the Allon
®
device is activated, a self-test must be
performed. Do not interrupt the self-test and wait until it is
finished.
NOTE:
During the Self-Test, the machine and display software
versions are shown.