DDT-063-000 Rev C Belmont Medical Technologies Page 15 of 106
Allon
®
Device
The Allon
®
device has a microprocessor that controls the water temperature
flowing into the ThermoWrap
®
worn by the patient. The algorithm to correct the
water temperature is based on the desired set point temperature and the
actual measured patient temperature (Core and Surface).
Water pressure and flow in the wrap is regulated by timed pauses of the flow
during clinical operation.
During the initial phase of regulation, the flow cycle is 12 minutes ON and 1
minute OFF.
In steady state (when the Core Temperature is within the Set Point range), the
cycle is 12 minutes ON and 12 minutes OFF.
The Allon
®
is equipped with a handle for easy transport.