DDT-063-000 Rev C Belmont Medical Technologies Page
79
of
106
Table 12: Allon
®
System Messages Troubleshooting Guide (Con’t)
Message
Cause of Problem
Action to be
taken
Comments
Check Surface Sensor
Misplacement of
surface temperature
probe in socket
Surface temperature
probe's adapter is
connected to the
Allon®
without the
temperature probe
Connect the
surface
temperature probe
to the appropriate
socket.
Connect
disposable
temperature
probe.
If the Surface temperature
probe was incorrectly
connected at power up with
the Core temperature probe
not connected, there will be
an alarm that can be muted
for 10 minutes.
If the Surface temperature
probe was incorrectly
connected during operation,
the alarm can be muted for
10 minutes.
Low Core Temperature
This message appears
when Set Point
Temperature is
<36°C and Core
Temperature is <32°C
or when the Core
Temperature is <28°C
The operator
needs to confirm
the location of the
core temperature
probe and touch
OK to continue.
When Core temperature is
below 32°C: The alarm can
be muted for 10 minutes.
When Core temperature is
below 28°C: The alarm can
be muted for 10 minutes.
Water temp. too low
When the water
temperature in the
system is less than
10°C (50°F)
Thermoregulation
stops.
Turn OFF the
system for 3
seconds and then
turn it ON again. If
the problem
persists, turn OFF
the Allon
®
and
contact your local
service
representative.
The alarm can be muted for
10 minutes.