Normothermia Mode
This is the default mode. During this mode, the system gets feedback from
both the patient’s and the water’s temperature, and adjusts the water
temperature accordingly, in order to reach and maintain the patient’s Set Point
temperature.
The default set point temperature is 37°C (98.6°F).
In this mode, the user can change the Set Point Temperature.
CAUTION!!
The desired set point temperature should only be
set by the physician or under the order of a
physician.
The Normothermia range is between 36°C and 38°C. Lower or
higher temperatures are indicated on the bar in
red
.
To change the Core Temperature Set Point:
1. Touch the temperature icon
. A temperature bar appears on the
screen.
2. Touch the arrow keys
/
or the bar scales on the screen in
order to change the set point temperature.
NOTE:
The
/
icons provide a change of 0.1°C. Each
bar scale provides a change of 1°C. The temperature
can be adjusted from 30 to 40°C.
3. Touch
OK
to confirm the selected temperature.
Figure 20: Select Set Point Temperature