![Belmont 200-00263 Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/belmont/200-00263/200-00263_user-manual_2730283071.webp)
Allon
®
User Manual
DDT-063-000 Rev C Belmont Medical Technologies
Page
71
of
106
Filter Replacement
The filter is for filtering hard soils or large particles and it is not intended for filtering
the water from bacterial contamination.
The filter must be replaced every twelve months.
NOTE:
The filter should be replaced only by Belmont Medical
Technologies authorized personnel / authorized biomedical
personnel. See the Service Manual for replacement
instructions.
NOTE:
Filter replacement could be performed by Belmont Medical
Technologies / biomedical authorized personnel if needed
more frequently than once a year (according to water
quality).
Annual checks should be performed every 12 months, as described in the Service
Manual.
System Check Service
The System Check service is initiated from the Services menu.
The System Check service performs a complete check of the system by checking
the functionality of the following components:
–
Screen and Buzzer
–
Pump
–
Wrap Connection
–
Pressure Meter
–
Heating and Cooling Unit
–
Temperature of Water Inflow and Water Outflow
Successful completion of the system check service indicates that the Allon®
device is operational.
NOTE:
If the Allon
®
was out of use for a long time, performing a full
System Check is recommended.
Содержание 200-00263
Страница 1: ...Allon User Manual DDT 063 000 Rev C English...
Страница 17: ...Allon User Manual DDT 063 000 Rev C Belmont Medical Technologies Page 17 of 106 Side View Figure 3 Side View...
Страница 18: ...Allon User Manual DDT 063 000 Rev C Belmont Medical Technologies Page 18 of 106 Rear Panel Figure 4 Rear View...