Section 2: Adjustable Alarm Limits
Adjustable alarm limits allow you to adjust the alarm limits which will trigger an
alarm in the system.
The adjustable alarms are:
–
High Patient Temperature
o
Range from 38°C to 40°C in 0.5°C steps
–
Low Patient Temperature
o
Range from 30°C to 35°C in 0.5°C steps
–
High Water Temperature
o
Range from 36°C to 42°C in 0.5°C steps
Figure 27: Adjustable Alarm Limits
NOTE:
Changing the alarm limits should occur only under order
of a physician.
NOTE:
Once the alarm limits are set, the limits remain fixed and
will not return to defaults.