-2-
Belfort AWOS Annual Maintenance Form, Jan 2013
BELFORT
INSTRUMENT
Proprietary Information
5.
Barometric Pressure Sensor (refer to Maintenance Manual section 5.2)
DCP S/N: Reference
Belfort AWOS
Acceptable
Adjustment
Value
_________ Standard(A)
Reading (B)
Tolerance
Required?
After Adjustment
BP1
_________
_________
±0.02 in Hg
Y ___ N ___
_________
BP2
_________
_________
±0.02 in Hg
Y ___ N ___
_________
Review the past year maintenance log to see if the DCP (with pressure sensors) had been replaced. If
it was replaced, indicate that fact here.
DCP Replaced
Yes
No
From the maintenance logs, review the previous measured pressure sensors values and document any
change in the measured sensor values since the last altimeter check.
Sensor Drift
BP1_________ BP2 _________ < +- .02 in Hg.
Record any configuration difference(s) identified _____________________________
6. Visibility Sensor (refer to Maintenance Manual section 5.4)
Review the past years maintenance log and verify that no replacement of the Visibility sensor was
made. If a replacement was made, verify that P/N and Revision installed is acceptable to current
authorized VDD baseline.
Yes
No
Record any configuration difference(s) identified _____________________________
7. Perform the Annual Wind Speed Sensor confidence validation (refer to Maintenance Manual section
5.5)
Pass
Fail
Review the past years maintenance log and verify that no replacement of the Wind Speed sensor was
made. If a replacement was made, verify that P/N and Revision installed is acceptable to current
authorized VDD baseline.
Yes
No
Record any configuration difference identified __________________________________
8. Perform the Annual Wind Direction Sensor confidence validation (refer to Maintenance Manual
section 5.6)
Pass
Fail
Review the past years maintenance log and verify that no replacement of the Wind Direction sensor
was made. If a replacement was made, verify that P/N and Revision installed is acceptable to current
authorized VDD baseline.
Yes
No
Record any configuration difference identified ______________________________
Содержание AWOS
Страница 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Страница 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Страница 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Страница 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Страница 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Страница 71: ...BELFORT INSTRUMENT Proprietary Information Appendix B Belfort AWOS Tri Annual Annual 5 Year Maintenance Plan ...
Страница 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Страница 76: ...BELFORT INSTRUMENT Proprietary Information Appendix D Standard Unicom Radio ICOM A110 VHF Radio Specifications ...
Страница 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...
Страница 93: ...BELFORT INSTRUMENT Proprietary Information Appendix G Belfort AWOS Atomic Clock Verification Procedure ...