-45-
BELFORT
INSTRUMENT
Proprietary Information
10.2
T
EMPERATURE
,
H
UMIDITY
/D
EW
P
OINT
S
ENSOR
R
EMOVAL
P
ROCEDURES
:
SAFETY PRECAUTIONS
1. Do not perform outdoor Belfort AWOS maintenance tasks during a thunderstorm.
2. Use extreme caution when lowering the tilt-over tower whenever ice or snow has accumulated
on the tower cross arm or the sensors. The extra weight may cause the tower to fall over
rapidly.
TOOLS REQUIRED
1.
Phillips head screwdriver
2.
Socket wrench and ¾" socket
Removal Procedures: Refer to Figure 6.
1.
At the DATU display terminal, issue a NOTAM advisory to alert users that the Belfort
AWOS system is being serviced prior to beginning investigation of this fault. The
NOTAM to be issued is "
Belfort AWOS system being serviced, do not use current
weather data outputs
." Verify that this NOTAM advisory is displayed on the operator
screen prior to beginning this maintenance tasks. Refer to the "Belfort AWOS Operator
Manual" for instructions on how to enter and delete NOTAMs.
2.
Remove tower-securing bolts and tilt the tower until the cross arm is at a comfortable
working height to access the Temperature/Humidity Sensor. Secure the tower in that
position by tying the attached tower hold-down rope to the base of the Belfort AWOS
tower as shown in Figure 2.
3.
Power down the power enclosure by opening the AC and DC circuit breakers.
4.
Loosen the jam nut on bottom of the plastic sensor adapter a few turns.
5.
Remove the sensor and sensor connector assembly through the bottom of the plastic
adapter.
6.
Disconnect the sensor probe from the sensor cable connector.
7.
Inspect the sensor cable connector for damage, corrosion or bent or broken pins. Replace
as necessary.
8.
Contact Belfort Instrument Co. customer support to make arrangements for replacement
of defective components.
9.
Refer to the Belfort AWOS Installation and Checkout manual Section 9.0 for installation
and checkout of the replacement parts.
10.
Document sensor replacement on the FAA form 6030-1
Содержание AWOS
Страница 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Страница 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Страница 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Страница 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Страница 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Страница 71: ...BELFORT INSTRUMENT Proprietary Information Appendix B Belfort AWOS Tri Annual Annual 5 Year Maintenance Plan ...
Страница 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Страница 76: ...BELFORT INSTRUMENT Proprietary Information Appendix D Standard Unicom Radio ICOM A110 VHF Radio Specifications ...
Страница 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...
Страница 93: ...BELFORT INSTRUMENT Proprietary Information Appendix G Belfort AWOS Atomic Clock Verification Procedure ...