-21-
BELFORT
INSTRUMENT
Proprietary Information
17.
Type “FC” and press enter. Make sure the Extco number in the configuration information for
Cal_ExtCo is the same extinction coefficient as on the calibration scatter plate. If not, enter
the correct extco number from the calibration scatter plate into the configuration data. Press
(ESC) key to save new value and get back to the terminal screen.
18.
Type “FS” and type “Y” at the “are you sure” query. When the routine has finished, the
terminal will display the new span calibration value and prompt for acceptance. Acceptable
values range between 0.5 and 4.0.
19.
If the span calibration value is out of range, repeat step 18. If out of range again, the instrument
should be considered faulty and taken out of service.
Follow the Belfort AWOS System
Component Failure Instructions in Section 5.10.
20.
Record the acceptable value displayed on the Tri-Annual maintenance data record form.
21.
Type “Y” and the new span value will be written to EEPROM. Remove the scatter plate and
store it for protection.
22.
Type “FR” to reset the sensor and exit the diagnostic mode.
23.
Disconnect the notebook computer and reconnect the sensor RS232 cable to the
communication cable running up to the DCP at the bottom of the tower. Apply Dow Corning
#4 liberally to the connection to protect against moisture.
24.
If the battery backup option is installed, switch the DC circuit breaker back on if battery backup
is present
25.
If additional maintenance tasks are to be performed, skip to next maintenance task. If
maintenance tasks are complete, toggle the “ALERT MESSAGE” switch 4 times to return the
unit to service.
26.
Verify the system begins outputting visibility information after a ten (10) minute algorithm
delay.
27.
If all maintenance tasks are complete, issue another NOTAM advising users that “
Belfort
AWOS
maintenance activities were completed, system has been returned to normal
operation
.” The expiration date and time of this NOTAM should coincide with the same
expiration date and time of the previous advisory NOTAM that was issued to put the system
in the maintenance mode. Verify that this NOTAM advisory is correctly displayed on the
operator screen. Refer to the
"Belfort AWOS Operator Manual"
for instructions on how to
enter and delete NOTAMs
.
28.
If you need additional assistance or have questions concerning these maintenance or calibration
tasks, contact Belfort Instrument Co. customer support at 1-800-937-2353.
Содержание AWOS
Страница 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Страница 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Страница 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Страница 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Страница 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Страница 71: ...BELFORT INSTRUMENT Proprietary Information Appendix B Belfort AWOS Tri Annual Annual 5 Year Maintenance Plan ...
Страница 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Страница 76: ...BELFORT INSTRUMENT Proprietary Information Appendix D Standard Unicom Radio ICOM A110 VHF Radio Specifications ...
Страница 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...
Страница 93: ...BELFORT INSTRUMENT Proprietary Information Appendix G Belfort AWOS Atomic Clock Verification Procedure ...