-9-
BELFORT
INSTRUMENT
Proprietary Information
6. Check the Advisor Status Log Page for any past or recurrent problems that have not been
addressed as current.
7. If doing a Tri-Annual or Annual performance test, document any unexplained recurrent error
flags on the Tri-annual or Annual “Facility Performance and Adjustment” data form. All
actions taken to correct errors identified will need to be documented and addressed on the FAA
Form 6030-1.
5.1.2 DATU
A
NNUAL
M
AINTENANCE
–
In addition to completing
scheduled tri-annual maintenance
tasks, DATU
annual maintenance consists of verifying operation of the Atomic Clock and defragmenting
the DATU computer hard drive and verification of the installed Belfort AWOS software configuration. It
is not necessary to stop the Belfort AWOS application during disk defragment.
To verify the Belfort AWOS Atomic clock, follow the Verification Procedure in Appendix G.
Follow the steps below to defragment the computer hard drive:
1.
Minimize the Advisor Window and any other running Belfort AWOS programs
2.
Empty the Recycle bin.
3.
Click on the “Start” button
4.
Click on “My Computer”
5.
Right click on “Local Disk, “C:”
6.
Select “Properties”
7.
Select the “Tools: tab
8.
Select and click on the “Disk Cleanup” button to perform that function.
9.
Click on the “Defragment Now” button
10.
Note the “% Free Space”
If “% Free Space is less than 15% or < 3G ,” notify Belfort Instrument Company for support.
11.
Click the “Defragment” button
12.
Once the defragmenting process is complete you may exit all windows opened in the above listed
steps.
13.
Maximize the Advisor window and any other previously minimized Belfort AWOS programs
Follow the steps below to check the version status of the installed Belfort AWOS software:
1.
Belfort AWOS Advisor:
a)
Select "Help" from the Advisor Menu.
b)
Select "About" from the submenu.
c)
The version number is located on the bottom of the window.
d)
Record the software version on the Annual Configuration Validation Performance and
Maintenance Record.
2.
Belfort AWOS Phone:
a)
Right click on the Belfort AWOS Phone icon located in the system tray.
b)
Select “Configure” from the popup menu.
c)
The version number is located on the top of the window.
d)
Record the software version on the Annual Configuration Validation Performance and
Maintenance Record.
Содержание AWOS
Страница 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Страница 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Страница 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Страница 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Страница 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Страница 71: ...BELFORT INSTRUMENT Proprietary Information Appendix B Belfort AWOS Tri Annual Annual 5 Year Maintenance Plan ...
Страница 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Страница 76: ...BELFORT INSTRUMENT Proprietary Information Appendix D Standard Unicom Radio ICOM A110 VHF Radio Specifications ...
Страница 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...
Страница 93: ...BELFORT INSTRUMENT Proprietary Information Appendix G Belfort AWOS Atomic Clock Verification Procedure ...