![Belfort AWOS Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/belfort/awos/awos_maintenance-manual_2725781035.webp)
-32-
BELFORT
INSTRUMENT
Proprietary Information
supply. If no AC short is found, then suspect the circuit breaker is faulty. Replace the
circuit breaker.
4.
Check inline fuses. If +14VDC is present and the inline fuse is OK, remove tower-securing
bolts and tilt the tower until the cross arm is at a comfortable working height to access the
wind Direction sensor. Secure the tower in that position by tying the attached tower hold
down rope to the base of the Belfort AWOS tower as shown in Figure 2. Put the Belfort
AWOS in “service mode” by toggling the “ALERT MESSAGE” toggle switch located on
the lower half of the hinged tower section six (6) times.
5.
Loosen the 4 screws securing the DCP to the tower and extend the DCP so the cable
attachments can be checked on the back of the DCP. Make sure the 4 wire power cable is
securely seated in the DCP.
If this does not return the AC power status indicator to service
and the DCP continues to run on battery backup, remove the 4 wire power cable from the
DCP. Note that this removes all power from the DCP. Measure the DC voltages across
the 4 pins of the power cable as follows:
Pin 1 + to 2- should m14.00 vdc from DCP power supply (PS)
Pin 4 + to 3- should m12 vdc from the battery
If the voltages are both present, then suspect the DCP AC status circuit is faulty.
When convenient, the DCP should be sent to the factory for repair or replaced.
If the PS DC voltage output is missing, then suspect a DC power supply failure, or loss of
AC input voltage to the power supply, or a DC voltage conductor in the tower power cable
is broken or shorted.
If the AC voltage in missing, then suspect an open circuit breaker, AC wiring short, AC
Mains supply feed is off.
6.
If the Voltage to the AC monitoring circuit is restored, re-attach the DCP to the tower with
it screws.
7.
Untie the rope and SLOWLY erect the tower to the vertical position and secure it by
installing and tighten-up (DO NOT OVER TIGHTEN) the three (3) tower-securing bolts.
NOTE: Be careful to avoid pinching the wires on the inside of the hinged Tower Section
when raising the tower.
8.
If additional maintenance tasks are to be performed, skip to next maintenance task. If
maintenance tasks are complete, toggle the “ALERT MESSAGE” switch 4 times to return
he unit to service.
9.
If a system component has been determined to be faulty or suspect bad, go to Section 5.10
and follow those instructions. Else, at the DATU display operator’s terminal, when all
maintenance actions are complete, issue another NOTAM advising users that “
Belfort
AWOS
maintenance activities were completed, system has been returned to normal
operation
.” The expiration date and time of this NOTAM should coincide with the same
expiration date and time of the previous advisory NOTAM that was issued to put the system
in the maintenance mode. Verify that this NOTAM advisory is correctly displayed on the
operator screen. Refer to the
"Belfort AWOS Operator Manual"
for instructions on how to
enter and delete NOTAMs
.
Battery Condition – Error flag is present if AC power is off and backup battery is low.
Содержание AWOS
Страница 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Страница 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Страница 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Страница 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Страница 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Страница 71: ...BELFORT INSTRUMENT Proprietary Information Appendix B Belfort AWOS Tri Annual Annual 5 Year Maintenance Plan ...
Страница 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Страница 76: ...BELFORT INSTRUMENT Proprietary Information Appendix D Standard Unicom Radio ICOM A110 VHF Radio Specifications ...
Страница 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...
Страница 93: ...BELFORT INSTRUMENT Proprietary Information Appendix G Belfort AWOS Atomic Clock Verification Procedure ...