PN B07493AD
4-2
Care and Maintenance
Instrument Care
4
If chamber icing occurs, defrost the system and wipe moisture out of the chamber before use.
a.
To defrost the system, set the temperature to 30°C for 20 minutes.
(These are suggested settings which may be adjusted as appropriate for your laboratory
conditions.)
NOTE
Before using any cleaning or decontamination methods except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not damage the
equipment.
Cleaning
Clean the centrifuge frequently. Always clean up spills when they occur to prevent corrosives or
contaminants from drying on component surfaces.
1
To prevent accumulations of sample, dust, and/or glass particles from broken sample tubes,
keep the chamber clean and dry by frequent wiping with a cloth or paper towel.
a.
For thorough cleaning, wash the chamber using a mild detergent such as
Beckman Solution 555 (339555), diluted 10 to 1 with water.
b.
Rinse thoroughly and dry completely.
2
Clean the centrifuge exterior surfaces by wiping with a cloth dampened with Solution 555,
diluted 10 to 1 with water.
a.
Do not use acetone.
3
Remove the rotor from the centrifuge and clean the drive hub regularly using Solution 555
(diluted 10 to 1 with water) and a soft brush.
a.
Rinse thoroughly and dry completely.
Tube Breakage
1
If a glass tube breaks, and all the glass is not contained in the bucket or rotor, be sure to
thoroughly clean the chamber.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......