![Beckman Coulter Allegra X-14 sERIES Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/beckman-coulter/allegra-x-14-series/allegra-x-14-series_instructions-for-use-manual_2719693033.webp)
PN B07493AD
2-9
Operation
Programmed Operation
2
Creating a New Program
1
Press
PROGRAM
, then use the keypad or arrow keys to select a program number that is not in
use, or one that you want to overwrite, and press
ENTER/SAVE
.
•
The program number flashes.
2
Enter run parameters (rotor, speed, time, temperature, acceleration/deceleration rate
settings). Press
ENTER/SAVE
.
•
If you entered an unacceptable parameter, an error message will be displayed.
a.
(Press
CE
and enter an acceptable value, then press the next run-parameter key.)
3
When all parameters are entered, press
ENTER/SAVE
.
•
The program is saved into memory and “
Program saved
” appears on the display.
•
The program number displayed will stop flashing.
•
If any parameters are then modified, the program number will show a dash.
4
Press
ENTER/SAVE
and then
START
to run the programmed run.
NOTE
If the
PROGRAM
key is pressed while the program number is flashing, program mode is exited
without saving entered parameters into program memory.
Recalling a Program
1
Press
PROGRAM
, then use the keypad or arrow keys to enter the number of the program.
a.
Press
ENTER/SAVE
to recall a program or press
PROGRAM
to exit program mode.
•
The program parameters are displayed if
ENTER/SAVE
was pressed.
2
Press
ENTER/SAVE
and
START
to begin the programmed run.
•
The run starts, the green light next to the
START
key flashes, and the rotor begins to spin.
NOTE
To begin a run,
ENTER/SAVE
must always be the last key pressed before pressing
START.
If you
wait more than 5 seconds, the
START
key will not activate. If this happens, press
ENTER/SAVE
and
START
again to begin the run.
•
The run will end automatically when the time display counts down to zero.
a.
(Press
STOP
to terminate a run at any time.)
•
The green light next to the
STOP
key will flash while the rotor decelerates.
•
A tone will sound when the rotor has come to a complete stop.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......