Operation
Preparation
2
5
Check that all parameters are correct and that the door is closed (door LED will be lit). Press
ENTER/SAVE
, then press
START
(within 5 seconds).
6
Wait for the run to end, or end the run by pressing
STOP
.
7
When the rotor stops (a tone sounds and the
DOOR
LED lights), press
DOOR
to open the chamber
door; lift the door.
Preparation
Prepare the rotor for centrifugation as described in the applicable rotor manual.
For runs at other
than room temperature, refrigerate or warm the rotor beforehand for fast equilibration
.
Installing the Rotor
The power must be turned on before you can unlock and open the chamber door. (To end a run for
any reason, do not turn the power switch off; press
STOP
instead.)
1
Turn the power switch on.
•
The control panel display shows “
Allegra X-14
” or “
Allegra X-14R
” and the software version.
2
Press
DOOR
. The centrifuge will accept this command only when the rotor is at rest and the
DOOR
LED is lit.
•
The door opens and the LED goes out.
3
Install the rotor according to directions in the rotor manual.
a.
Ensure that the rotor is seated on the drive hub.
b.
Avoid bumping the control panel keys during rotor installation or removal.
4
Close the chamber door by pressing down firmly on both sides.
•
The door locks.
•
When the latches engage, the LED next to the
DOOR
key lights up.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......