PN B07493AD
2-1
CHAPTER 2
Operation
Introduction
This section contains centrifuge operating procedures. A summary is provided at the start of this section.
If you are an experienced user of this centrifuge, you can turn to the summary for a quick review of
operating steps. Refer to the applicable rotor manual for instructions on preparing the rotor for
centrifugation.
WARNING
Handle body fluids with care because they can transmit disease. No known test
offers complete assurance that they are free of micro-organisms. Some of the
most virulent—Hepatitis (B and C) and HIV (I–V) viruses, atypical mycobacteria,
and certain systemic fungi—further emphasize the need for aerosol protection.
Handle other infectious samples according to good laboratory procedures and
methods to prevent spread of disease. Because spills may generate aerosols,
observe proper safety precautions for aerosol containment. Do not run toxic,
pathogenic, or radioactive materials in this centrifuge without taking appropriate
safety precautions. Biosafe containment should be used when Risk Group II
materials (as identified in the World Health Organization
Laboratory Biosafety
Manual)
are handled; materials of a higher group require more than one level
of protection.
Do not use the centrifuge in the vicinity of flammable liquids or vapors, and do not
run such materials in the instrument. Do not lean on the centrifuge or place items
on it while it is operating.
Summary of Run Procedures
For runs at other than room temperature, refrigerate or warm the rotor beforehand for fast equilibration.
Manual Run
1
Turn the power switch on (
I
).
2
Press
DOOR
to open the chamber door; lift the door open.
3
Install the rotor according to the applicable rotor manual, then close the chamber door.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......