Operation
Manual Operation
2
2
Use the keypad to enter required speed.
•
Entered speed appears on the speed display.
•
(If the entered speed is out of the speed range of the installed rotor, an error message will
appear and the input will be cleared.
a.
Enter an acceptable value from the range displayed.)
3
Press the next run-parameter key or
ENTER/SAVE
.
•
The rotor will accelerate or decelerate to the new speed.
•
You can change the set speed at any time during a run by repeating steps 1 through 3.
Entering Run Time
Run time can be set for up to 99 hours and 59 minutes or for a hold (continuous) run. The time
display begins counting down when the rotor starts to spin. For a timed run, the run automatically
terminates when the set time reaches zero and deceleration begins at the selected rate; a tone
sounds when the rotor has stopped. For a hold run, the elapsed time is displayed and the run
continues until
STOP
is pressed.
1
Press
TIME
.
•
If
HOLD
is displayed, press
TIME
again to enter time.
2
Use the keypad or arrow keys to enter required time.
•
Entered time appears on the display.
•
If you entered a number higher than 59 minutes, the centrifuge automatically recalculates
the time in hours and minutes.
or
a.
Press
TIME
again for a hold run.
•
HOLD
appears on the display.
3
Press the next run-parameter key or
ENTER/SAVE
.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......