![Beckman Coulter Allegra X-14 sERIES Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/beckman-coulter/allegra-x-14-series/allegra-x-14-series_instructions-for-use-manual_2719693037.webp)
PN B07493AD
3-3
Troubleshooting
User Messages
3
R1
and
R2
–
Rotor ID problem, call
Service
No magnets
identified, or magnets
incorrectly identified
Run shuts down
with maximum
brake
•
Make sure the rotor in use is a compatible
Beckman Coulter rotor (see
Available
Rotors
in
CHAPTER 1
). Check that magnets
are in place on rotor bottom. Press
CE
to
clear the message.
•
Call Beckman Coulter Field Service.
R3
–
Rotor ID mismatch,
check rotor
The entered rotor
number is not the
same as the rotor
identified
Run shuts down
with maximum
brake
Press
CE
to clear message. Enter the correct
rotor entry identification.
R4
and
R5
–
No rotor match, see
manual
The system cannot
identify the rotor
Run shuts down
with maximum
brake
•
Make sure the rotor in use is a compatible
Beckman Coulter rotor (see
Available
Rotors
in
CHAPTER 1
). Check that magnets
are in place on rotor bottom. Press
CE
to
clear the message.
•
Call Beckman Coulter Field Service.
S1
–
System error, call
Service
There is a problem
with the system
control software,
EPROM, or RAM
System shuts down,
no brake
Call Beckman Coulter Field Service.
S8
–
System error, door
locks for 45 minutes,
see manual
There is a problem
with the system
control software
Run shuts down
with no brake, door
locks for 45 minutes
Wait 45 minutes then press
CE
to clear
message, restart run. If problem persists, call
Beckman Coulter Field Service.
H1
,
H2
, and
H4
–
Speed error, call
Service
Speed control
problem
Run shuts down
with maximum
brake.
Call Beckman Coulter Field Service.
H3
,
H5
, and
H6
–
Speed error, door
locks for 45 minutes,
call Service
Speed control
problem
Run shuts down
with no brake; door
locks for 45 minutes
Call Beckman Coulter Field Service.
Table 3.1
Diagnostic Message Chart
(Continued)
Diagnostic
Number/Message
Problem
Result
Recommended Action
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......