PN B07493AD
3-2
Troubleshooting
User Messages
Table 3.1
Diagnostic Message Chart
Diagnostic
Number/Message
Problem
Result
Recommended Action
P1
–
Power failure
occurred, see manual
System or facility
power failure
Run shuts down
Press
CE
to clear message. Restart run.
Remaining time shows on display.
L1
through
L6
–
Door latch, call
Service
Latches are not
operating properly
Error message
appears; run shuts
down with no brake
Press down on both sides of the door until the
DOOR
LED comes on. If you close the door
repeatedly and the problem continues, call
Beckman Coulter Field Service.
a
L7
–
Can’t open door, see
manual
Door latches do not
release
Door doesn’t open
when
DOOR
key is
pressed
Press down on the door, then press
DOOR
again. If door still doesn’t open call
Beckman Coulter Field Service. (Procedures for
retrieving your sample are at the end of this
section.) If the door opens but the message
stays on the display, press
CE
to clear.
C1
–
Rotor temp exceeds
4C above set
Rotor temperature
exceeds temperature
setting by more than
4°C but less than 8°C
after equilibration
Run continues; the
message is
informational
Press
CE
to clear message.
C2
–
Rotor temp exceeds
8C above set, call
Service
Rotor temperature
exceeds temperature
setting by more than
8°C after equilibration
Run shuts down
Call Beckman Coulter Field Service.
C3
and
C5
–
Temp error, call
Service
Refrigeration system
error
Run shuts down
with maximum
brake
Call Beckman Coulter Field Service.
T1
and
T2
–
Temp, door locks for
45 minutes, call
Service
System temperature
problem
Run shuts down
with no brake; door
locks for 45 minutes
Call Beckman Coulter Field Service.
T3
–
Temp, door locks for
45 minutes, see
manual
Motor overheated
Run shuts down
with no brake; door
locks for 45 minutes
Wait 45 minutes for motor to cool down, then
restart run. If problem persists, call
Beckman Coulter Field Service.
D1
through
D9
–
Drive error, call
Service
Drive system problem Run shuts down
with no brake
Call Beckman Coulter Field Service. Before
trying to open the door, listen carefully and
make sure that no sound is coming from the
chamber (indicating a spinning rotor). Follow
the directions under
Retrieving Your Sample in
Case of Power Failure
, below.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......