PN B07493AD
2-2
Operation
Summary of Run Procedures
4
Press
ROTOR
, then press
ROTOR
again or use the arrow keys to select the rotor.
5
Press
RPM/RCF
, then use the keypad to enter the run speed. Press
RPM/RCF
a second time to
toggle between RPM and RCF modes.
6
Press
TIME
, then use the keypad to enter the run time (to 99 hrs, 59 minutes); or press
TIME
twice to toggle between hold (continuous) run and set time mode.
7
Press
TEMP °C
, then use the keypad to enter the required run temperature 2 to +40°C.
8
Press
ACCEL
, then use the keypad or arrow keys to enter the selected acceleration rate number,
1
(
SLOW
) through
10
(
MAX
).
9
Press
DECEL
, then use the keypad or arrow keys to enter the selected deceleration rate number,
0
(
OFF
) through
10
(
MAX
).
10
Check that all parameters are correct and that the door is closed (door LED will be lit). Press
ENTER/SAVE,
then press
START
(within 5 seconds).
11
Wait for the set time to count down to zero, or end the run by pressing
STOP
.
12
When the rotor stops (a tone sounds and the
DOOR
LED lights), press
DOOR
to open the chamber
door; lift the door.
Programmed Run
1
Turn the power switch on (
I
).
2
Press
DOOR
to open the chamber door.
3
Install the rotor according to the applicable rotor manual, then close the chamber door.
4
Press
PROGRAM
, use the keypad or arrow keys to enter the required program number, then
press
ENTER/SAVE
.
Содержание Allegra X-14 sERIES
Страница 4: ...PN B07493AD iv Revision History...
Страница 8: ...PN B07493AD viii Safety Notice Chemical and Biological Safety...
Страница 24: ...PN B07493AD 1 10 Description Available Rotors...
Страница 50: ...PN B07493AD Warranty 2 Beckman Coulter Inc Allegra X 14 Series Centrifuge Warranty...
Страница 53: ......