PN B07493AD
1-9
Description
Available Rotors
1
Available Rotors
The following Beckman Coulter rotors can be used in the Allegra X-14 series centrifuges. The rotors
are described in individual manuals that accompany each rotor.
Rotor Profile
Description
Rotor
Selection
Code
Max
RPM
a
Max RCF
b
(
g
)
at
r
max
Number
of Tubes
Nominal
Capacity
Rotor
Manual
Number
FX6100 Fixed Angle
r
max
= 98.0 mm
FX6100
10,200
11,419
6
100 mL
GX-TB-005
SX4750 Swinging Bucket
Tube-and-bottle buckets,
r
max
= 207.8 mm
Multiwell-plate carriers,
r
max
= 183.2 mm
SX4750
SX4750
4,300
4,000
4,303
3,283
4
750 mL
4
96 mL
GX-TB-003
SX4750A Swinging
Bucket (ARIES)
Tube-and-bottle buckets,
r
max
= 207.8 mm
Multiwell-plate carriers,
r
max
= 183.2 mm
SX4750A
SX4750a
4,300
4,000
4,303
3,283
4
750 mL
4
96 mL
GX-TB-004
a. Maximum speeds are based on a solution density of 1.2 g/mL. At upper temperature and humidity ambient conditions, swinging bucket rotor
speed may require reduction.
b. Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and speed (
r
2
) to the standard acceleration of
gravity (
g
) according to the following formula: RCF =
r
2
/
g
— where
r
is the radius in millimeters,
is the angular velocity in radians per second
(2
RPM/60), and
g
is the standard acceleration of gravity (9807 mm/s
2
). After substitution: RCF = 1.12
r
(RPM/1000)
2
.
Содержание Allegra X-14 sERIES