15
Cartridge packs will come bagged with four manifold
connection caps.
Part
Application
DIamond Kit Organic Free R/O & Distilled Feed
D50280 Ultra-Low Organics, Type 1 Water,
Reverse Osmosis or Distilled
Water Feed
DIamond Kit Organic Free Deionized Feed
D50281 Ultra-Low Organics, Type 1 Water,
Deionized Water Feed
DIamond Kit Type 1 R/O & Distilled Feed
D50282 Low Organics, Type 1 Water,
Reverse Osmosis or Distilled
Water Feed
DIamond Kit Type 1 Deionized Feed
D50283 Low Organics, Type 1 Water, DI
DI Water Feed
Each cartridge pack includes one 0.2 micron absolute
final filter.
Cartridge Pack Installation
1.
Open front (left side) door. Turn the cartridge
pack so the caps are facing you. Remove the
caps.
2.
Verify that each of the four posts on the car-
tridge pack has an o-ring.
3.
Lift unit manifold upwards, insert cartridge pack
and align the pack so that it mates with the unit
manifold.
4.
Lower unit manifold until it is flush with the top
of the cartridge pack.
5.
Hand tighten wing head screws securely.
6.
Close door.
Initial Operation
Note
Do not install the 0.2 micron filter and
bell assembly at this time.
Caution
Do not allow the NANOpure DIamond
Analytical to operate unless water is
available to the unit.
Warning
Depressurize system prior to removing
cartridge pack.
Warning
Allow cartridge pack to warm to room
temperature before installation.
Manifold