18
13. Press the UP or DOWN arrow to select (YES)
and press ENTER. Display will now read, “Unit
Sanitizing.”
14. Unit may now be left unattended until display
reads, “Sanitization Complete: Press ENTER.”
15. After sanitization is complete the unit will return
to the (Idle) mode.
16. Press the START/STOP switch to return the unit
to the normal recirculation mode.
17. Remove the new 0.2 micron filter and bell
assembly from the bag and insert it into the dis-
penser arm manifold. Gently turn it clockwise
until it is fully seated.
18. Remove the protective cap from the filter bell.
Turn the dispense knob to the left or right and
flush 1-2 liters of water through the 0.2 micron
filter. The product water is now ready for use.
I
NITIAL
O
PERATION
Note
For more demanding applications
where low TOC water is required, a
rinse of 15-20 liters through the car-
tridges and 0.2 micron filter may be
necessary.
0.2 Micron