41
Fuse Replacement
1.
Turn off the NANOpure DIamond Analytical and
disconnect it from the power supply by removing
the power cord directly below the main power
switch on right side of unit. Locate the fuse
drawer above the power cord receptacle.
2.
Pull out the fuse drawer located in the power
entry module.
3.
Remove old fuses and replace with
fuses of the same type and rating. (See
“Replacement Parts.”)
4.
Replace fuse drawer.
5.
Reattach the power cord.
6.
Operate normally.
Shutdown
If the NANOpure DIamond Analytical is to be shut down
for an extended period of time, the unit should be com-
pletely drained and the cartridge pack removed to prevent
the growth of bacteria. If the system has remained inac-
tive and full of water for more than 96 hours, the unit
should be drained, sanitized and new cartridge pack
installed prior to use.
M
AINTENANCE AND
S
ERVICING
Warning
Replace fuses with those of the same
type and rating.