Dimensions and Clearance Requirements
Dimensions
13.4” W x 19.5” H x 16.9” D (34.0 x 49.5 x 42.8 cm)
Clearances
Sides - 9” (22.9 cm) minimum for servicing.
Above - 3” (7.6 cm) minimum for removal of the top cover.
Front - 4.75” (12.1 cm) minimum for opening the front door.
Feed Water Requirements
Types
RO, DI, distilled.
TOC
Less than 1.0 ppm.
Turbidity
1.0 N.T.U. maximum.
Pressure Range
Gravity feed to 100 psig (7kg/cm2) maximum.
Temperature Range
4°C - 40°C (40-104°F)
TDS (CaCO3)
< 70 ppm
Silica
< 1 ppm
Silt
< 5% SDI
Product Water
Water Quality
Resistivity
> 18.0 m
Ω
-cm
TOC
Less than 10 ppb
Flow Rate
1.5 lpm maximum at minimum inlet feed water pressure 10 PSIG with a new final filter.
Volumetric Dispense
Accuracy: ± 5%
Repeatability: ± 3%
Electrical Requirements
The NANOpure DIamond Analytical is equipped with 2 power cords to be plugged into an electrical outlet of
the appropriate voltage.
Voltage and Frequency (Nominal)
Model D11901
100-240 VAC, 47-63 Hz, 1 phase
6