23
Use of Standby Mode
At the end of the work day, place the NANOpure
DIamond Analytical in Standby mode for the night. To
place the unit in Standby mode:
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx
xx/xx/xx,” press STANDBY.
—OR—
2.
From the xx.x M
Ω
-cm or x.xx uS/cm operational
display, Press STANDBY.
After STANDBY is pressed the display will read “Standby”
and the green LED will slowly blink. For 10 minutes of
every hour while the pump is energized, the display will
read “Standby (Recirculating) xx minutes.” Press
START/STOP to return to the recirculating mode.
Selecting the Set Point
The NANOpure DIamond Analytical electronics include a
user programmable set point which alerts you when water
quality falls below the programmed set point. The set
point is user selectable from 1-18 M
Ω
-cm (factory set at
10 M
Ω
-cm.).
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx
xx/xx/xx,” press the UP or DOWN arrow until
display reads, “Advanced Menu?”
2.
Press ENTER.
3.
Proceed through the options by pressing
ENTER until “Adjust Set Point” is displayed.
4.
Press the UP or DOWN arrow to choose “Yes.”
5.
Press ENTER.
6.
Press the UP or DOWN arrow until the desired
value is displayed.
U
SER
S
ETTINGS
Note
When in Standby mode, press
START/STOP to continue normal
operation.