System Sanitization
After sanitization is complete (approximately 3.5 hours),
place the unit into its normal (recirculation) mode to allow
the system to recirculate overnight after cleaning.
1.
From the (Idle) mode, press the UP or DOWN
arrow until the display reads, “Sanitize Unit?”
2.
Press ENTER.
3.
Display will prompt, “Disconnected Inlet Water
(NO).”
4.
Disconnect feed water supply at the quick-dis-
connect inlet valve.
5.
Press the UP or DOWN ARROW to select (YES)
and then press ENTER. Display will now
prompt, “Injected Sanitant (NO)?”
6.
Remove solution syringe from wrapper.
7.
Open front (left side) door. Remove luer cap on
system injection port, located on the right side
of the cartridge pack manifold by turning coun-
terclockwise.
8.
Remove the syringe luer cap and attach the
syringe to the system luer fitting.
9.
Slowly inject the solution into the system and
remove syringe. Avoid injecting air into system.
10. Replace luer cap on injection port. Close door.
11.
Press the UP or DOWN arrow to select (YES)
and then press ENTER. Display will now read,
“Reconnected Water (NO).”
12. Attach feed water supply removed in step 4.
13. Press the UP or DOWN arrow to select (YES)
and press ENTER. Display will now read, “Unit
Sanitizing.”
37
M
AINTENANCE AND
S
ERVICING
Note
The sanitization timer will display after
six months, reminding you to sanitize
the unit. Complete the system saniti-
zation procedure and reset the saniti-
zation timer according to the
“Resetting the Sanitization Timer” sec-
tion in this manual.
Warning
Disconnect from the power supply
prior to maintenance and servicing.
Warning
Avoid splashing disinfecting solution
on clothing or skin.
Ensure all piping connections are tight
to avoid leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety
instructions and material safety data
sheets.
This device to be used with water
feeds only.
Sanitizing/cleaning agents must be
used in compliance with instructions in
this manual. Failure to comply with the
above could result in explosion and
personal injury.
Warning
Refer servicing to qualified personnel.