17
System Sanitization Procedure
Your NANOpure DIamond Analytical has been shipped
with a cleaning syringe (Cat. No. CMX25). When the sys-
tem is new or whenever a new cartridge pack is installed,
the system should be sanitized (cleaned). It is best to per-
form sanitizing procedures when the system or water is
not required. The entire sanitization procedure will take
approximately 3.5 hours to complete. Sanitize your
NANOpure DIamond as follows:
1.
From the (Idle) display, press the UP or DOWN
arrow until display reads, “Sanitize Unit?”
2.
Press ENTER.
3.
Unit display will ask if you have “Disconnected
Inlet Water (NO).”
4.
Disconnect feed water supply at the quick-dis-
connect inlet valve on left side of unit.
5.
Press the UP or DOWN ARROW to select (YES)
and then press ENTER. Display will now read,
“Injected Sanitant (NO).”
6.
Open front (left side door) to expose cartridge
pack. Remove solution syringe from wrapper.
7.
Remove luer cap on system injection port,
located on the right side of the cartridge pack
manifold by turning counterclockwise.
8.
Remove syringe luer cap and attach syringe to
the system luer fitting.
9.
Slowly inject solution into system, and remove
syringe. Avoid injecting air into the system.
10. Replace luer cap on injection port and close
door.
11.
Press the UP or DOWN arrow to select (YES)
and then press ENTER. Display will now read,
“Reconnected Water (NO).”
12. Attach feed water supply removed in step 4.
I
NITIAL
O
PERATION
Warning
Avoid splashing disinfecting solution
on clothing or skin.
Ensure all piping connections are tight
to avoid leakage.
Ensure adequate ventilation.
Carefully follow manufacturer’s safety
instructions and material safety data
sheets.
This device to be used with water
feeds only.
Sanitizing/cleaning agents must be
used in compliance with instructions in
this manual. Failure to comply with the
above could result in explosion and
personal injury.
Note
After sanitization is complete, the sani-
tization timer will automatically be
reset for approximately six months.
Injection Port