12
Water Connections
Feed Water Connection
1.
Locate the length of 3/8” (.95 cm) O.D. tubing
provided with the unit with a quick disconnect
insert on one end and a 3/8” (.95 cm) O.D. X
1/4” (.64 cm) NPT tubing adapter on the other.
2.
Install the tubing adapter onto your
incoming water line. We recommend a
customer supplied shut off valve be
installed in your feed water line. Do not
connect the feed water to your NANOpure
DIamond Analytical. You will connect the feed-
water during the “Initial Sanitization” procedure
explained later in this manual. Cartridge pack
must be installed beofre connecting water line.
3.
If a pressure switch is to be used, see the
“Installing a Float or Pressure Switch” section
for more information.
Atmospheric Drain Connection
When the NANOpure DIamond Analytical flushes, the
water used is sent to drain through this connection. To
install:
1.
Locate the drain water tubing. This is the 1/4”
(.64 cm) O.D. tubing that is approximately 6
ft.(1.9 m.) long. The atmospheric drain fitting is
located on the lower left side of the NANOpure
DIamond Analytical.
2.
Route the other end of the drain water tubing to
an atmospherically vented drain and make a
connection. Ensure there are no kinks in the
tubing and that it proceeds in a downward
plane. Proceed to “Initial Operation.”
I
NSTALLATION
Water Connections - Left side view
Caution
Do not connect feed water until direct-
ed to do so during “Initial Sanitization.”
Pump
Interlock
Remote 2
Remote 1
Feed
Water Inlet
Atmospheric
Drain