6. Final Adjustment and Configuration
Barco – OverView cDG67-DL – R5976624– installation manual – Revision 03 – August-2005
______________________________________________________________
6-25
How to determine the DIS Calibration Factor
The DIS Calibration Factor is determined by CAST.
The following description applies if there is urgent need to define the DIS Calibration Factor, and if no CAST is
available.
To determine the DIS Calibratin Factor, it is mandatory to set the DIS correction factor to
0.0.
This has to be done in the menu Color/Brightness| Set Brightness
To calibrate the DIS, a lux meter is required.
Press the sensor of the lux meter on the screen.
Read the value of the lux meter
Select
Service | Color/Brightness | Set Brightness
Figure 6-34
DIS value
Make sure that
Absolute
is enabled
Read the
DIS
value.
Now the
DIS Calibration Factor
can be calculated:
Brightness
Brightness
or
ation Fact
DIS Calibr
DIS Value
Luxmeter
=