English–DPI 610/615 IS Instruction Manual | 19
Transmitter Simulator (TX SIM) Task
4.8 Transmitter Simulator (TX SIM) Task
When used with an external voltage source (see Section
5.6), provides a current output
proportional to the calibrator’s measured output pressure (indicated pressure on indicator only
version). Select task
Tx
SIM
. Press
EXIT
to skip set-up screen if parameters are correct.
On completion of
Tx
SIM
set-up, the display is configured as follows:
Connect an external power source to the output loop as detailed in Section
5.6.
To subsequently change any of the
Tx
SIM
scaling parameters, press
CHANGE VALUE
key (F1)
to obtain the TX Simulation set-up display.
To change the pressure units, press
INPUT
and select the required scale units. If the required
scale units are not listed, press
SETUP
, select
SETTINGS
and proceed as detailed in
Tx SIMULATION
SET VALUE
SET VALUE
SET VALUE
SET VALUE
LOWER POINT
UPPER POINT
CHANGE
VALUE
Tx SIMULATION
LOWER POINT
UPPER POINT
CHANGE
VALUE
Tx SIMULATION
LOWER POINT
UPPER POINT
CHANGE
VALUE
Tx SIMULATION
LOWER POINT
UPPER POINT
CHANGE
VALUE
TASK : TX SIM
SNAPSHOT MODE
PRESSURE INT
CHANGE
VALUE
CURRENT
OK
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......