Maximum Safe Input Parameters:
Installation
•
Provide additional protection for equipment that may be damaged in service.
•
Installation should be carried out by qualified plant installation technicians in compliance with
the latest issue of EN
60079-14.
Declaration Requirements - EU Directive 2014/34/EU
The DPI
610
IS and DPI
615
IS pressure calibrators are designed and manufactured to meet the
essential health and safety requirements not covered by the EU Type Examination Certificate
BAS02ATEX1174X when installed as detailed above.
lout (SK6)
RS232
External Transducer
U
o
= 7.9 V dc
U
o
= 7.6 V dc
U
o
= 7.9 V dc
C
i
= 0
I
o
= 82 mA dc
I
o
= 155 mA dc
L
i
= 0.1 mH
P
o
= 162 mW
P
o
= 0.43 W
C
i
= 0
C
i
= 0.15 µF
L
i
= 0
L
i
= 0.9 mH
U
m
= 250 V
C
o
= 8.6 µF
L
o
= 0.3 mH
SK1, SK2, SK3 and SK6
U
i
= 30 V
l
i
= 100 mA
P
i
= 1.0 W
WARNING
Do not use tools on the pressure calibrator that might cause
incendive sparks - this can cause an explosion.
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......