Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 43
Voltage Input Range (5 V)
2. Select 5 V range for calibration:
3. Link the voltage input terminals with a short lead and enter the zero point:
4. Remove the shorting link from the voltmeter terminals, connect the Vin terminals of the
instrument to a variable voltage source and connect a digital voltmeter across the source:
Figure
8
-
1: 5 V Input Range Calibration Circuit
Vs Voltage source
SELECT CAL OPTION
SELECT CAL OPTION
SELECT RANGE
30 VOLTS
5 VOLTS
CALIBRATION
VOLTAGE IN
CURRENT IN
CURRENT OUT
PRESSURE INT
TEMPERATURE
Apply nominal zero value
Press ENTER key when
applied value is stable
CALIBRATING VOLTAGE
RANGE +/- 5V
CALIBRATING VOLTAGE
RANGE +/- 5V
Enter Applied Value:
CHANGE
VALUE
RUN
CHANGE VALUE
Store zero point?
YES
NO
CALIBRATING VOLTAGE
RANGE +/- 5V
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
V
Vs
+
-
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......