Copyright 2007 Baker Hughes Company.
English–DPI 610/615 IS Instruction Manual | 45
Voltage Input Range (30 V)
4. Remove the shorting link from the voltmeter terminals, connect the Vin terminals of the
instrument to a variable voltage source and connect a digital voltmeter across the source:
Figure
8
-
2: 30 V Input Range Calibration Circuit
5. Set the variable supply voltage to 30 V ± 0.1 V and enter the measured full-scale voltage:
6. Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to quit the
CALIBRATION
and
SETUP
modes.
7. Verify the instrument calibration by applying the test voltages given in Table
to the voltmeter (after both voltage ranges have been calibrated).
8. Disconnect the calibration/test equipment.
Vs Voltage source
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
V
Vs
+
-
SET VALUE
Apply nominal FS value
Press ENTER key when
applied value is stable
CALIBRATING VOLTAGE
RANGE +/- 30V
CALIBRATING VOLTAGE
RANGE +/- 30V
Enter Applied Value:
CHANGE
VALUE
RUN
CALIBRATING VOLTAGE
RANGE +/- 30V
Enter Applied Value:
CHANGE
VALUE
RUN
CALIBRATING VOLTAGE
RANGE +/- 30V
Store FS point?
YES
NO
Calibration Complete
Press ENTER to continue
Applied
Measured
CALIBRATING VOLTAGE
RANGE +/- 30V
RESULTS
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......