English–DPI 610/615 IS Instruction Manual | 9
Typical Calibration Set-up (Pressure to Voltage)
2.5 Typical Calibration Set-up (Pressure to Voltage)
Connect a device under test to the instrument as shown below:
2.5.1 General Procedure
Use the hand-pump to pressurize the system to the required level as indicated on the display.
Allow the display to settle and screw the volume adjuster in or out as a fine adjustment to the
required pressure. Record the input:
Voltage
, reading at each applied pressure.
2.6 Zero Display Reading
Both the input and output readings can be set to zero by using the
ZERO
key and if the displayed
reading is within 5% of zero. To zero either the
INPUT
or
OUTPUT
displays, proceed as follows:
A
External pressure source (indicator instruments only)
B
Pressure regulator
C
Pressure to voltage device
D
Barrier
E
Excitation 10 V
F
Safe area
PRESSURE INT
VOLTAGE
CURRENT
PRESSURE
UNITS
TASK : BASIC
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......