Page 2-2
Bacharach, Inc.
Instructions 67-9319
INSTALLATION
INJECTOR CALIBRATOR CD3
Figure 2-4. Electrical Connections
Figure 2-5. Water Connections
2.3 Utility Connections
Electrical
1. First, check the nameplate on the calibrator for
electrical data. The plate is located behind the
door on the left side of the calibrator to the left
(on top of frame angle) beneath the motor.
Refer to Section "Technical Data" for information
on required utilities.
2. Check that the On/Off Switch is OFF. Remove
the lower left hand side panel covering the
electrical box.
3. Wire the proper electrical power line to the
motor starter in the electrical control box after
punching appropriate hole.
CAUTION
CALIBRATOR CAN BE DAMAGED. If
there is no fluid in the reservoir at this time,
do not allow Clamp (circulating) pump or
main drive motor to run more than a few
seconds to check rotation.
4. After wiring, jog the main drive motor (turn On/
Off Switch ON and OFF quickly) to check that
the direction of rotation is the same as shown
by the arrow on the motor.
5. If the motor direction is wrong, switch any two
wires and check again. The motor must rotate
counter clockwise as viewed from the drive end.
Water
6. Connect a cold water line to the fitting at the
rear of the calibrator marked WATER IN (1/4
inch NPT).
7. Connect a drain line to the fitting marked
WATER OUT (1/4 inch NPT).
NOTE:
Do not install any valves, shutoffs
or other restrictions on drain line. Water
goes in under line pressure and must be free
to run out.
1 Nameplate
4 Motor Rotation Direction
2 On/Off Switch
8 Air In
2 On/Off Switch
7 Water Out
6 Water In
3 Electric Power Inlet
Содержание 67-7622
Страница 26: ...Page 2 8 Bacharach Inc Instructions 67 9319 INSTALLATION INJECTOR CALIBRATOR CD3 NOTES ...
Страница 54: ...Page 3 28 Bacharach Inc Instructions 67 9319 OPERATION INJECTOR CALIBRATOR CD3 NOTES ...
Страница 68: ...Page 4 14 Bacharach Inc Instructions 67 9319 MAINTENANCE AND TROUBLESHOOTING INJECTOR CALIBRATOR CD3 NOTES ...
Страница 96: ...Page A 2 Bacharach Inc Instructions 67 9319 APPENDIX INJECTOR CALIBRATOR CD3 NOTES ...