Chapter 2 Measurement Operation
AE-4022
2-18
4
Prepare test strips
a
Take out the required number of test strips from the bottle.
b
Close the test strip bottle cap securely.
5
Start normal measurement
a
Confirm that the icon (
) for normal measurement appears
on the [Standby screen].
• When using the “Auto start” mode, go to step
.
• When using the “Cycle start” mode, press
.
REFERENCE:
The currently set test strip type is displayed on the [Standby
screen]. To change the test strip type, please refer to “3.7.5.
No.002: Test strip type” on page 3-22.
IMPORTANT:
After taking the test strips out of the bottle, immediately
close the cap or the test strips in the bottle will absorb
moisture or dirt in the air, which may deteriorate reagents
and make the test strips useless.
IMPORTANT:
Different procedures at this stage are required, depending
on the setting of the “operational mode when turning ON”
parameter. (See “
Operational mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the
“operational mode when turning ON” parameter, see “3.7.8.
No.005: Operational mode when turning ON” on page 3-27.
REFERENCE:
When the buzzer setting is ON, the buzzer sounds to help maintain the proper timing of the dipping process. The
dipping duration is key for accurate results.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......