Chapter 2 Measurement Operation
AE-4022
2-6
2.3
Preparation for measurement
Before starting measurement, check the waste box, feeder, and thermal recording paper.
2.3.1
Check before measurement
1
Check for used test strips
a
Pull out the waste box to see if it is filled with used test strips.
If the waste box is full of test strips, discard them.
2
Open the maintenance cover
a
Press and hold the two cover open buttons on either side of the
instrument, (
(1)
in the right figure) to open the maintenance cover
(
(2)
in the right figure).
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Discard used test strips in accordance with local regulations for biohazardous waste.
NOTE:
The special test strips for the AUTION ELEVEN
do not
come with the instrument. Please purchase a sufficient
supply before starting measurement.
Waste box
Cover open
button
Motor
(1)
(1)
(2)
NOTE:
When the maintenance cover is opened, the power is
automatically turned OFF.
Do not
touch the motor, which may be hot.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......